dbConn Object ( [strHost:dbConn:private] => localhost:3306 [strUser] => everfine [strPwd] => everfine2020 [strDbName] => everfine )

Products

Single Phase Digital Power Meter -

PF2010 Digital Power Meter  (Multifunction & high accuracy & wide bandwidth)
Name

PF2010 Digital Power Meter (Multifunction & high accuracy & wide bandwidth)

  • Description
  • Parameter
  • Application
There are kinds of measurement functions for PF2010 Digital Power Meter. It can realize tests in a wide range of frequency from 10Hz to 500kHz and DC. And the accuracy of fundamental frequency measurement is 0.06%. It is with multiple functions, wide frequency and high accuracy which is very suitable for high accuracy testing situation.
Back

Items

Voltage

C urrent

Measurement range

0.1V600V

2.5mA20A(direct input)

Input range (RMS)

 

30/60/100/200/400/600V

Direct input:0.5/1/2/5/10/20A

External input:50/100/200mV

Range switch

Manual / Automatic

Test mode

RMS / MEAN / DC

Frequency range

DC/AC10Hz500kHz

Measurement accuracy

DC:

±(0.04%Reading + 0.08%Range+1 digit)

±(0.08%Reading + 0.16%Range+1 digit)

10Hz45Hz

±(0.03%Reading + 0.05%Range+1digit)

±(0.05%Reading + 0.05%Range+1digit)

45Hz65Hz

±(0.03%Reading + 0.03%Range+1digit)

±(0.04%Reading + 0.04%Range+1digit)

65Hz250Hz

±(0.05%Reading + 0.1%Range+1digit

±(0.05%Reading + 0.2%Range+1digit)

250Hz1kHz

±(0.05%Reading + 0.5%Range+1digit

±(0.05%Reading + 1%Range+1digit)

1kHz10kHz

±(0.05%Reading + 1% Range+1 digit)

±(0.05%Reading + 2%Range+1 digit)

10kHz100kHz

±(0.01×(f-10)%Reading + 1.5%Range+1 digit)

±(0.02×(f-10)%Reading + 2.5%Range+1 digit)

100kHz500kHz

±(0.03×(f-100)%Reading + 3%Range+digit)

±(0.15×(f-100)%Reading + 5%Range+digit)